Abstract

A fine tube type glucose sensor, which has a sensing region at the inside wall of fine tube tip, was prepared and applied for minimally invasive glucose monitoring. The enzyme glucose oxidase was immobilized on the platinum surface located in the tube by the combination of electrodeposition and electropolymerization of o-phenylenediamine. Amperometric responses of the prepared electrodes to glucose were measured at a potential of 0.60 V (vs Ag/AgCl). The tube type electrode showed good response with good linear relationship up to the glucose concentration of 22.4 mM, which was actually higher than general sensor without outer film for permeability restriction. This may be due to it unique sensor structure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.