Abstract
A capacitor–resistor combined bank to calibrate both the capacitance ratio and dissipation factor over the whole range for commercial capacitance and tan $\delta $ measuring bridges was fabricated. A fabricated secondary working standard for capacitance ratio calibration was realized by connecting double-shielded capacitors in parallel employing a substitution and build-up technique, while a secondary working standard for dissipation factor calibration was attained by connecting shielded capacitors in parallel with a shielded T-resistor network. The design of the capacitor–resistor bank and calibration results for the two commercial bridges are presented and compared with specifications provided by manufacturers including the measurement uncertainty.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Instrumentation and Measurement
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.