Abstract

A novel iodide ion-selective electrode has been produced based on a molecular deposition technique in which water-soluble porphyrin was alternatively deposited with water-soluble polypyrrole on a 2-aminoethanethiol modified silver electrode. The potentiometric response is independent of pH of the solution between pH 1 and 7, while it is dependent on the nature of the medium. The electrode has a linear dynamic range between 1.6×10 −6 and 0.1 M with a Nernstian slope of 59 mV/decade and a detection limit of 1.0×10 −6 M in acetate buffer (0.1 M, pH 4.6). The electrode has the advantages of low resistance, short conditioning time and fast response.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.