Abstract

We investigated the size and mode dependencies of the resonance frequencies and quality factors of PMMA (polymethyl methacrylate) and PC (polycarbonate) cantilevers under atmospheric pressure in order to obtain fundamental data for polymer-based VOC (volatile organic compound) sensors. PMMA cantilevers and PC cantilevers are fabricated by hot embossing, bonding, and polishing techniques. Variations in the vibration mode result in large variations in the resonance frequencies and quality factors of these cantilevers. The measured resonance frequencies correlated well with the theoretical undamped resonance frequency of a flexural-vibrating cantilever. The measured quality factors varied from 10 to more than 100, depending on the vibration mode. A higher mode tended to exhibit a higher quality factor.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.