Abstract

Flow studies have been conducted, in vitro, on prototype microfabricated blood flow sensors. The envisioned measurement principle is the detection of the rotation of a micromachined polysilicon rotor of diameter 300 mu m. Repeatable rotation rates as a function of media velocity were obtained for both nitrogen and water. Fluid flow asymmetries, necessary to produce a torque on the rotor, were created by integrating a 2 mu m thick polysilicon cap over half the rotor. Blade deflection, due to intrinsic stress in the polysilicon films, was eliminated by thermal annealing. This allows the rotor, 2 mu m thick, to rotate in the 7 mu m gap between the substrate and the polysilicon cap. Operation of the device in heparinized dog blood resulted in considerable numbers of erythrocytes sticking to the polysilicon elements.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.