Abstract

At the LOTUS facility, an extremely efficient on-line detector system, based on the detection of the charged particles associated with the [sup 6]Li(n,[alpha])t reaction, has been designed, fabricated, and tested. The system offers an interesting possibility for directly measuring the tritium production rate (TPR) at any experimental site. The charged particles emitted in opposite directions can be detected by a double parallel plate ionization chamber (DIC) configuration. The real events are identified by employing a coincidence circuit. The complete fabrication details, testing under different conditions, measurement of TPR, and its comparison with the liquid-scintillation method (LSM) are detailed. The DIC response to thermal neutrons agrees well with the theoretical calculations. Also, the detector system is insensitive to a contact gamma dose rate of 1.3 rem/h. The direct TPR measurements and the salient feature of higher efficiency in comparison with the LSM are demonstrated. The TPR determined by both methods are in excellent agreement. 14 refs., 11 figs., 1 tab.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.