Abstract

Raja Ramanna Centre for Advanced Technology (RRCAT) has developed a 2K Cryostat for characterization of Superconducting RF (SRF) cavities, under Indian Institution Fermilab Collaboration (IIFC). A vertical Test Stand (VTS) is a facility for qualifying bare SRF cavities for their required performance by measuring quality factor and cavity accelerating gradient at a cryogenic temperature of 2K. The VTS cryostat has been designed for a large testing aperture for testing variety of SRF cavities including 325 MHz Spoke resonators, 650 MHz and 1.3 GHz multi-cell SRF cavities. The engineering design and analysis of VTS cryostat has been carried out using ASME B&PV Code and Finite Element Analysis. Fabrication of Cryostat was carried out in strict accordance with ASME B&PV code under joint supervision of engineers from RRCAT and Fermilab. The paper describes the fabrication & acceptance testing aspects of the VTS cryostat at the vendor location and at RRCAT.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.