Abstract

Some of the problems of developing a two-dimensional (2-D) transducer array for medical imaging are examined. The fabrication of a 2-D array material consisting of lead zirconate titanate (PZT) elements separated by epoxy is discussed. Ultrasound pulses and transmitted radiation patterns from individual elements in the arrays are measured. A diffraction theory for the continuous wave pressure field of a 2-D array element is generalized to include both electrical and acoustical cross-coupling between elements. This theory can be fit to model the measured radiation patterns of 2-D array elements, giving an indication of the level of cross-coupling in the array, and the velocity of the acoustic cross-coupling wave. Improvements in bandwidth and cross-coupling resulting from the inclusion of a front acoustic matching layer are demonstrated, and the effects of including a lossy backing material on the array are discussed. A broadband electrical matching network is described, and pulse-echo waveforms and insertion loss from a 2-D array element are measured.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call