Abstract
This study uses simple and reliable microfabrication techniques to fabricate CE biochips, integrating a novel contactless conductivity detector in a miniaturized detection system in a microfluidic biochip. The off-channel electrodes are deposited around side channels by Au sputtering and patterned using a standard "lift-off" process. A vacuum fusion bonding process is employed to seal the lower substrate containing the microchannels and the electrodes to an upper glass cover plate. The variations in the capacitance between the semicircular detection electrodes in the side channels are measured as different samples and ions pass through the detection region of the CE separation channel. Samples of Rhodamine B, commercial sports drinks, mineral waters, and a red wine, respectively, are mixed in different buffer solutions, separated, and successfully detected using the developed device. The semicircular detection electrodes for the contactless conductivity detector have microscale dimensions and provide a valuable contribution to the realization of the lab-on-a-chip concept.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.