Abstract
A PZT actuator integrated onto cantilever structure was fabricated for high speed atomic force microscopy (AFM). Five different electrodes were used to investigate the effect of top electrodes on the adhesion and the electrical properties in the PZT capacitors. The PZT capacitors with RuO2 top electrodes exhibited the best characteristics of five electrodes. The tip deflection and the resonant frequency of the PZT actuator were 11 μim at 10 V and 79 kHz, respectively. The PZT actuator provided much better AFM image quality and imaging speed than those done by using the conventional bulk PZT tube scanner. The creep distortion in the AFM image was greatly improved by using the high speed PZT film actuator.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.