Abstract

A continuously regenerated cationic impurity removal device (CR-CRD) has been fabricated and applied for ion chromatography (IC). The removal of cationic impurities is realized by electrodialytically replacing the cationic impurities with hydronium ions. The device is configured in a sandwich structure and the central eluent channel is respectively isolated from both electrodes by stacked cation exchange membranes and a bipolar membrane (BPM) plus stacked anion exchange membranes. The eluent channel is packed with cation exchange resins in hydronium form and their continuous regeneration can be achieved by electrodialysis. A desirable feature of the device is gas-free, and no degasser is required. It showed sufficient ability to remove cationic impurities, as indicated by > 99.9 % removal of 10 mL of 1 mM LiOH solution injected (∼10 μmol) or continuous removal of 1 mM LiOH solution at the flow rate of 1 mL/min (1 μmol/min). A useful application was for sample pretreatment in nuclear power industry, by eliminating strong matrix interference of the sample containing LiOH (1 mM) and boric acid (2000 mg/L) with trace anion analysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.