Abstract
BackgroundHolding endoscopes by hand when performing eye surgery reduces the dexterity of the surgeon.MethodsA robotic endoscope holder called “Eye Explorer” is proposed to hold the endoscope and free the surgeon's hand.ResultsThis device satisfies the engineering and clinical requirements of eye surgery. The force for manual operation is less than 0.5 N. The observable ranges inside the patient's eye considering horizontal and vertical perspectives are 118° and 97°, and the motion of the holder does not interfere with the surgeon's hand and other surgical devices. The self‐weight compensation can prevent the endoscope from falling when extra supporting force is released. When comparing the external force exerted on the eye by the Eye Explorer with that in case of manual operation, a decrease of more than 15% can be observed. Moreover, the consumption time of endoscope view adjustment using the Eye Explorer and manual operation does not significantly differ.ConclusionThe Eye Explorer allows dual‐hand operation, facilitating a successful endoscopic eye surgery.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: The international journal of medical robotics + computer assisted surgery : MRCAS
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.