A new process, based on the micro-co-extrusion of preceramic precursors, has been studied for manufacturing ceramic microelectrodes to be used in biomedical applications. Commercially available silicon polymers were applied and proper doping resulted in electrically conductive ceramic filaments. Chemical reticulation and high-temperature pyrolysis were applied to convert the polymeric resins into Si-O-C ceramic materials. Circular microelectrodes were manufactured with diameters between 100 microm and 5 mm with a different number of inner conductive lines (from 1 to 80). The flexural strength of the filaments depended on the outer diameter size; doping with carbon black produced filaments with an average conductivity of approximately 0.4 S/cm for a 50% weight carbon black load. The results achieved by in vitro studies confirmed a good biological performance of Si-O-C ceramic structures with both hard and soft tissue cell models.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.