Abstract

A new cluster time-of-flight secondary ion mass spectrometry (TOF-SIMS) was developed using a size-selected gas cluster ion as a projectile. Since a large gas cluster ion can generate many low-energy constituent atoms in a collision with the surface, it causes multiple and ultra low-energy sputtering. The mean kinetic energy of constituent atoms is provided by dividing the acceleration energy of the gas cluster ion by the number of constituent atoms. Therefore, the sputtering can be controlled to minimize the decomposition of sample molecules and substrate material by precisely adjusting the number of constituent atoms (the cluster size) and/or acceleration energy of the gas cluster ion. The cluster size was selected on the basis of the time-of-flight method using two ion deflectors attached along the ion-beam line. A high resolution of 11.7 was achieved for the cluster size/size width ( M/Δ M) of Ar-cluster ions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.