Abstract

Abstract. The current limitation in the accuracy and precision of inter-element analysis in secondary ion mass spectrometry (SIMS) is the ability to find measurable quantities that allow relative differences in ionization and transmission efficiency of secondary ions to be normalized. In uranium–thorium–lead geochronology, the ability to make these corrections, or "calibrate" the data, results in an accuracy limit of approximately 1 %. This study looks at the ionization of uranium and thorium oxide species, which are traditionally used in U–Pb calibration, to explore the conditions under which isotopologues, or molecular species whose composition differs only in the isotopic composition of one or more atoms in the molecule, remain in or deviate from equilibrium. Isotopologue deficits of up to 0.2 (200 ‰) below ideal mixing are observed in UO2+ species during SIMS gechronological analyses using the SHRIMP IIe SIMS instrument. These are identified by bombarding natural U-bearing minerals with an 18O2− primary beam. The large anomalies are associated with repeat analyses down a single SIMS sputtering crater (Compston et al., 1984), analysis of high-uranium, radiation-damaged zircon, and analysis of baddeleyite. Analysis of zircon under routine conditions yield UO2+ isotopologue anomalies generally within a few percent of equilibrium. The conditions under which the isotopologue anomalies are observed are also conditions in which the UOx-based corrections, or calibration, for relative U vs. Pb ionization efficiencies fail. The existence of these isotopologue anomalies suggest that failure of the various UOx species to equilibrate with each other is the reason that none of them will successfully correct the U / Pb ratio. No simple isotopologue-based correction is apparent. However, isotopologue disequilibrium appears to be a more sensitive tool for detecting high-U calibration breakdowns than Raman spectroscopy, which showed sharper peaks for ∼ 37 Ma high-uranium zircons than for reference zircons OG1 and Temora. U–Th–Sm / He ages were determined for aliquots of reference zircons OG1 (755±71 Ma) and Temora (323±43 Ma), suggesting that the broader Raman lines for the Temora reference zircons may be due to something other than accumulated radiation damage. Isotopologue abundances for UO+ and ThO+ and their energy spectra are consistent with most or all molecular species being the product of atomic recombination when the primary beam impact energy is greater than 5.7 keV. This, in addition to the large UO2+ instrumentally generated isotopologue disequilibria, suggests that any attempts to use SIMS to detect naturally occurring isotopologue deviations could be tricky.

Highlights

  • Determining the timing of geologic events is a fundamental constraint for unravelling the history of our planet

  • One of the major advances in this field was the invention of the SHRIMP (Sensitive High Resolution Ion MicroProbe), which has been used for U–Pb geochronology of zircon for the last 34 years (Froude et al, 1983)

  • 1050 cm−1 1050.09 1052.85 1055.13 1052.6 1055.29 1054.35 1050.28 1053.61 1052.8 1050.9 1054.64 1052.65 www.geosci-instrum-method-data-syst.net/6/523/2017/. It appears that UO2 isotopologues are in or close to equilibrium in those circumstances where the secondary ion mass spectrometry (SIMS) UO+x / U+based calibration successfully works in the SHRIMP

Read more

Summary

Introduction

Determining the timing of geologic events is a fundamental constraint for unravelling the history of our planet. In the 120 years since the discovery of radioactivity, the use of radioactive decay has been an increasingly versatile, accurate, and precise way of measuring geologic time. One of the major advances in this field was the invention of the SHRIMP (Sensitive High Resolution Ion MicroProbe), which has been used for U–Pb geochronology of zircon for the last 34 years (Froude et al, 1983). Magee Jr. et al.: Isotopologue disequilibrium in SIMS

Methods
Results
Discussion
Conclusion