Abstract

Magnetic particle imaging exploits the non-linear magnetization of superparamagnetic iron-oxide particles to generate a tomographic image in a defined field-of-view. For reconstruction of the particle distribution, a time-consuming calibration step is required, in which system matrices get measured using a robot. To achieve artifact-free images, system matrices need to cover not only the field-of-view but also a larger area around it. Especially for large measurements- inevitable for future clinical application- this leads to long calibration time and high consumption of persistent memory. In this work, we analyze the signal in the outer part of the system matrix and motivate the usage of extrapolation methods to computationally expand the system matrix after restricting the calibration to the field-of-view. We propose a suitable extrapolation method and show its applicability on measured 2D and 3D data. In doing so, we achieve a considerable reduction of calibration time and consumption of persistent memory while preserving an artifact-free result.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.