Abstract

Data from dental extraoral and cone beam computed tomography (CBCT) exposures in Portugal (2019) were gathered, and patient doses for standard adult exams were evaluated. In panoramic X-rays, 442units (34% of the existing licensed units) were tested, with a third quartile value (PKA) of 82mGy.cm2. For cephalometric radiography (88units), the third quartile value (Ki) was 0.3mGy for the posteroanterior projection and 0.2mGy for lateral projection. In CBCT (69units), the doses for the placement of an upper first molar implant were evaluated with a third quartile value (PKA) of 820mGy.cm2. Due to the wide range of values (74-3687mGy.cm2), the CBCT data were divided by FOV dimensions: for small FOV (average FOV of 7×8 cm), a value of 580mGy.cm2 was obtained and for medium FOV (average FOV of 13×12 cm) a value of 1167mGy.cm2. The number of annual panoramic X-rays made in Portugal was 208 per 1000 inhabitants, which is higher than the value for other countries.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.