Abstract

An extraction system has been developed, essentially consisting of a pump, a rotary valve and a spectrophotometer. The sample, 12–25 μl, is introduced via the rotary valve into an aqueous stream (flow injection). The aqueous stream, containing the sample plug, is divided into small segments by an organic phase and led into a Teflon coil so that a regular pattern of the two phases is obtained. No air bubbles should be present. Separation of the two phases is achieved in a specially constructed fitting and the absorbance of the organic phase is measured. The construction and performance of the system are illustrated by analysis of caffeine samples. Up to 100 samples/h can be analysed with a relative precision of better than 1%.