Abstract

A series of new neutral hydrogen ion carriers suitable for application in H+-selective microelectrodes is presented. One carrier (ETH 1907) proves to be superior to tridodecylamine currently very much in use. Microelectrodes based on ETH 1907 in an optimized membrane composition exhibit a linear dynamic response function from pH 2 to 9 extended into the acidic range, a response time less than or equal to 5 s, and a resistance of about 35 G omega for a tip diameter of about 1 micron. This makes the electrode suitable for measurements at normal physiological intracellular pH as well as in acid physiological media. Measurements using this microelectrode in proximal tubule cells of isolated perfused frog kidney are presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.