Abstract

Matrix-assisted laser desorption/ionization time-of-flight mass spectrometry( MALDI-TOF MS) is a bacterial typing tool that was approved as a medical device in 2011. However, external accuracy control examination of bacterial typing using mass spectrometry is still only performed on a small scale. In this study, E. faecium and S. maltophilia were selected and tested according to established procedures using Score Values at 228 institutions. The Score Values for MALDI Biotyper were 2.43±0.08 for E. faecium and 2.38±0.08 for S. maltophilia; and those for VITEK MS/PRIME were 99.9±0.0 for E. faecium and S. maltophilia. These results suggest that it is useful to evaluate external accuracy control with Score Values using the procedures we have developed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.