Abstract
Many sources of uncertainty exist in the delivery of the prescribed dose to the specified target volume in external beam radiotherapy. Their magnitudes are compared with recommendations for the overall accuracy of dose delivery (+/- 5%). Five dose intercomparisons are reviewed, including a recent UK survey undertaken by the Institute of Physical Sciences in Medicine (IPSM). For phantom doses, standard deviations in the ratio of stated to delivered doses range from 1-3.5%, with larger uncertainties when irradiation of volumes within anthropomorphic phantoms is considered. In addition, uncertainties arise from geometrical discrepancies, and the role of portal films in their assessment is also considered. Imaging during radiotherapy will be increasingly important in the development of complex conformal treatments, where the shape of the target volume may be irregular.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.