Abstract

A summary is given of the theory of measurement of complex relative permeability and permittivity of low conductivity materials by the thin-sample method using a slotted transmission line. The equations from which to calculate the electrical properties from the measured parameters are extended to apply to a commercial slotted line and thus to a more complex system than that used by other workers. Some errors of the measurements are discussed and various correctional equations derived by other workers are given.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.