Abstract
At the boron neutron capture therapy (BNCT) facility in Petten, the Netherlands, 10B concentrations in biological materials are measured with the prompt gamma ray analyses facility that is calibrated using certified 10B solutions ranging from 0 to 210 ppm. For this study, newly certified 10B solutions ranging up to 1972 ppm are added. MCNP simulations of the setup range to 5000 ppm. A second order polynomial (as already used) will fit 10B-concentrations less than 300 ppm. Above 300 ppm a fitted third order polynomial is needed to describe the calibration curve accurately.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.