Abstract

The aim of this study was to establish traceable number concentration measurements of airborne particles beyond 10 μm in particle size. To this end, the primary standards for particle number concentration at the National Metrology Institutes of Switzerland and Japan were further developed to extend their measurement capabilities. Details on the upgraded setup are provided. An inter-comparison of the two primary standards using an optical particle counter as transfer standard showed that these agree well within the stated uncertainties at polystyrene (PS) equivalent optical diameter of 15 µm. Subsequently, four Model 3321 (TSI Inc., USA) aerodynamic particle size spectrometers (APS) were calibrated against the primary standard of Switzerland using size-certified PS spheres with optical/aerodynamic diameter up to 20 µm as test aerosols. The counting efficiency profile and unit-to-unit variability of the APS units were determined. The results presented here can be useful for the analysis and interpretation of data collected by the different atmospheric aerosol networks worldwide. The outlined methodology can also be applied in the calibration of automated bio-aerosol monitors.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call