Mass spectrometry-based immunopeptidomics enables the comprehensive identification of major histocompatibility complex (MHC) peptides from a cell culture as well as from tissue or tumor samples and is applied for the identification of tumor-specific and viral T-cell epitopes. Although mass spectrometry is generally considered an "unbiased" method for MHC peptide identification, the physicochemical properties of MHC peptides can greatly influence their detectability. Here, we demonstrate that highly hydrophobic peptides are lost during sample preparation when C18 solid-phase extraction (SPE) is used for separating MHC peptides from proteins. To overcome this limitation, we established an optimized protocol involving restricted access material (RAM). Compared to C18-SPE, RAM-SPE improved the overall MHC peptide recovery and extended the landscape of mass spectrometry-detectable MHC peptides toward more hydrophobic peptides.