Abstract

To the Editor: Sporadic cases of dislodgment of the occluders have been reported in the majority of artificial cardiac valvular prostheses. We report a case of expulsion of the disk of a Björk-Shiley aortic prosthesis. A 42-year-old woman was diagnosed by cardiac catheterization as having severe mitral stenosis and grade-3/4 aortic regurgitation. The patient underwent mitral commissurotomy and replacement of the aortic valve with a No. 21 Björk-Shiley tilting-disk prosthesis. The prosthesis was fixed in place with 16 single sutures (Ethiflex). In the immediate postoperative period the clinical picture and the phonocardiogram suggested significant aortic regurgitation. Repeat surgery was performed 13 days after the first procedure. It was discovered that an excessively large space existing between two of the sutures was responsible for the leak. In order to gain better access to this zone, we decided to remove the disk; we also rotated the valve within the sewing ring with the valvular holder designed by the manufacturer for this purpose. The leak was closed with a U suture, the disk was put back into position, and it was left functioning normally, being apparently well held in place. The patient was discharged from the hospital ten days later, without any evidence of aortic regurgitation. Seventeen days after the second operation, the patient was awakened from her sleep by a nightmare, and seconds later she lost consciousness. She was readmitted to our hospital in cardiac arrest. All manipulations to induce recovery failed, and the patient died. The autopsy revealed that the disk had dislodged from the rest of the prosthesis and had come to rest in the abdominal aorta. On examination of the valve, an excessive clearance between the struts and the disk was discovered. We attribute the detachment of the disk in our patient to an excessive clearance between the struts and the disk produced during the temporary removal of the disk. It is less likely that the deformation of the struts occurred during the rotation of the prosthesis with the valvular holder. One of the advantages of the Björk-Shiley prosthesis is the possibility of rotating the valve within the sewing ring after its implantation. We have performed this operation numerous times without any type of complication. The temporary removal of the disk of a Björk-Shiley aortic prosthesis was performed by Björk1Björk VO Byrd CL Yahr CL Greenberg JJ Long-term results of “simple thrombectomy” for thrombosed Björk-Shiley aortic valve prostheses.Ann Thorac Surg. 1975; 20 (in discussion): 272Google Scholar, 2Björk VO Henze A Encapsulation of the Björk-Shiley aortic disc valve prosthesis caused by the lack of anticoagulant treatment.Scand J Thorac Cardiovasc Surg. 1973; 7: 17-20Crossref PubMed Scopus (42) Google Scholar in his first two cases of thrombotic encapsulation of an aortic prosthesis; however, in his later cases, total thrombectomy was performed with a sucker after rotating the valve.1Björk VO Byrd CL Yahr CL Greenberg JJ Long-term results of “simple thrombectomy” for thrombosed Björk-Shiley aortic valve prostheses.Ann Thorac Surg. 1975; 20 (in discussion): 272Google Scholar This was our first case of dislodgment of the disk of a Björk-Shiley aortic prosthesis after 284 aortic valvular replacements with this type of prosthesis. It was the first time that we have temporarily removed the disk, and it will be the last.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.