Abstract

Slit radiography is a highly efficient means of suppressing scatter. However, when employed with a pulsating x-ray source, a grid line type of artifact results if the movement of the slit is not properly coupled to the periodicity of the source. Taking the exposure of the film to be the convolution of the slit transmittance and the radiation intensity, the conditions for a uniform exposure have been derived for the realistic configuration of a multiple-slit array and a slit transmittance that includes the focal spot and slit geometry. Expressions for the amplitude and spatial frequency of the artifact when the conditions are not met are derived. Methods of suppressing the artifact when its vanishing conditions are not met are also discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.