Abstract

In South African breast care centres, full-field digital mammography units provide breast imaging services to symptomatic and asymptomatic women simultaneously. This study evaluated the technical exposure parameters of 800 mammograms of which 100 mammograms had obvious mass lesions in the fibroglandular tissue. The average breast compression force of mammograms with mass lesions in the fibroglandular tissue was 18.4% less than the average breast compression force of mammograms without mass lesions. The average mean glandular dose (MGD), tube potential (kVp) and compressed breast thickness (CBT) values were 2.14 mGy, 30.5 kVp and 63.9 mm, respectively, for mammograms with mass lesions, and 1.45 mGy, 29.6 kVp and 56.9 mm, respectively, for mammograms without mass lesions. Overall, the average MGD and mean CBT of mammograms with mass lesion were significantly higher compared to those without mass lesions (p < 0.05), although there was no significant difference in their tube potentials (p > 0.05).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.