This paper discusses a methodology which uses a specially developed computer program that enables estimation of output parameters of an x-ray machine from a single test exposure. Exposure at skin entrance values can be estimated, as well as using the data from the program for organ dose estimates. To perform the estimation procedure, curves and analytical expressions for the curves are presented for conversion of half-value layers to total filtration and for conventional mR/mAs output curves. Results are given showing that the estimation procedure is reasonably accurate.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.