Abstract

Polyp miss-rate is a drawback of colonoscopy that increases significantly for small polyps. We explored the efficacy of an automatic computer-vision method for polyp detection. Our method relies on a model that defines polyp boundaries as valleys of image intensity. Valley information is integrated into energy maps that represent the likelihood of the presence of a polyp. In 24 videos containing polyps from routine colonoscopies, all polyps were detected in at least one frame. The mean of the maximum values on the energy map was higher for frames with polyps than without (P < 0.001). Performance improved in high quality frames (AUC = 0.79 [95 %CI 0.70 - 0.87] vs. 0.75 [95 %CI 0.66 - 0.83]). With 3.75 set as the maximum threshold value, sensitivity and specificity for the detection of polyps were 70.4 % (95 %CI 60.3 % - 80.8 %) and 72.4 % (95 %CI 61.6 % - 84.6 %), respectively. Energy maps performed well for colonic polyp detection, indicating their potential applicability in clinical practice.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.