Abstract

Exudate and its viscosity are critical in wound healing. Changes in viscosity can interfere with dressings properties as well as affect the diffusion of immune cells, nutrients, oxygen and bacteria. Current international standards for laboratory testing of wound dressings use a single low-viscosity solution, named as 'Test Solution A', which fails to simulate the diverse range of exudates encountered clinically. This study employs rheological analysis to characterise exudates viscosity, comparing cattle-derived samples to the test solution A. Results reveal non-Newtonian, shear-thinning behaviour in exudates, contrasting with the Newtonian behaviour of the test solution A. Although clinically classified as 'seropurulent', three exudate samples analysed at 37°C present with different viscosity at various shear rates, ranging from 30.8 (±14.7) to 6.5 (±1.9) mPas. Findings show that the current tests on dressings employing Test Solution A are missing the complexity of real exudates.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.