Abstract

A 2 3 factorial design and a CCD (Central Composite Design) were employed to optimize the operational conditions for determining Al, Ba, Ca, Cd, Cr, Cu, K, Mg, Na, Ni, Pb, Sn and Zn in soft drink samples by inductively coupled plasma optical emission spectrometry (ICP OES). Satisfactory parameters of merit were obtained (Plasma stability, linearity, SBR, BEC, LOD and LOQ). The accuracy was evaluated by means of a recovery study. Thirty five soft drink samples were analyzed. The results obtained were used in an exploratory analysis, in which principal component analysis and cluster analysis were applied in order to differentiate the several soft drink classes. Adequate LOQ values were obtained for all elements (0.34 μg L − 1 for Ca to 88.7 μg L − 1 for Mg) and the recovery study results ranged between 80 and 100%. Cluster analysis revealed four sample classes characterized by flavor and other characteristics, such as being light, diet, low calorie or regular beverages. A model with 4 PCs was obtained by means of a principal component analysis, which explained 99.8% of the total variance.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.