Abstract

Electrocorticography (ECoG) is an important area of research for Brain-Computer Interface (BCI) development. ECoG, along with some other biopotentials, has spectral characteristics that can be exploited for more optimal front-end performance than is achievable with conventional techniques. This paper optimizes noise performance of such a system and discusses an equalization technique that reduces the analog-to-digital converter (ADC) dynamic range requirements and eliminates the need for a variable gain amplifier (VGA). We demonstrate a fabricated prototype in 1p9m 65 nm CMOS that takes advantage of the presented findings to achieve high-fidelity, full-spectrum ECoG recording. It requires 1.08 μW over a 150 Hz bandwidth for the entire analog front end and only 7 bits of ADC resolution.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.