Abstract

A thermal model for predicting time-dependent temperature distributions during ultrasound heating was tested quantitatively. The relevant thermal processes incorporated in the model are heat conduction and ultrasound power absorption, and the required input parameters include the absolute ultrasound power, the shape and frequency of the ultrasound transducer, and the thermal and acoustical properties of the medium. Testing was done by heating an ultrasonically tissue-mimicking phantom with a 525 kHz, single element, focused source. The phantom has muscle-like acoustical properties and contains an array of copper-constantan thermocouples. Time-dependent temperature changes predicted with the thermal model were in very good agreement with those measured in the phantom, verifying the validity of the model for use in optimizing an ultrasound applicator design for a specific treatment situation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.