In a previous paper, one of the authors suggested an analytical method for calculation of the response function of an alpha spectrometer for the case of large solid angles. This paper describes the experimental verification of the method. Spectra of a well-known natural uranium sample were measured with a 450 mm 2 Si detector and compared to the theoretical predictions. The measurements were carried out with two different geometrical configurations. In both cases a good agreement was observed between experimental and theoretical results.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.