Abstract

A multi-square slot surface coil was developed for rodent MRI at 11.7 T. The coil prototype was 20 mm in diameter with six square slots. Electromagnetic simulations of the slotted and circular coils were conducted to evaluate the signal-to-noise ratio (SNR) and absorption rate (SAR). Experimental SNR results were also calculated and compared, and the comparison showed good agreement between experimental and numerical data. The slotted coil produces lower predicted SAR for points near the coil plane when compared to the SAR numerical values for the circular coil. The performance of prototype, measured via the Q factor, was $$Q_l 6.5 = Q_u$$ , and its noise figure was 1.6. A commercial circular coil was used to acquire phantom images for comparison purposes. The slotted surface coil showed comparable performance with a commercial circular coil in UHF MRI applications. These experimental and numerical results demonstrate the feasibility of the slotted coil design at 11.7 T.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.