Abstract
Gas-washing and water injection are two commonly used methods in generating certain moisture content in gas flow stream, but the control accuracy of these methods remains uncertain. The purpose of the study is to investigate the characteristic of water vapor generation of the two methods. The temperature and humidity recorder and FTIR gas analyzer have been adopted to monitor the moisture content in tests. It has been revealed that, in gas-washing process, the moisture content in gas stream is well below saturation, and long time is required before moisture content stabilized after water bath temperature or carrier gas flow rate changed. Water injection pump could directly control the amount of water injected into gas stream. However, the moisture content generated in this way is not stable but oscillates in cycles, whose frequency and magnitude are closely related to water injection amount.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.