Abstract

This study aimed to explore the accuracy of non-invasive temperature measurement based on proton resonance frequency (PRF) phase subtraction in microwave ablation (MWA). The signal change of the agar phantom during the ablation process was monitored by the gradient echo sequence under 1.5 T superconducting magnetic resonance imagining (MRI), and the temperature change was converted by the phase subtraction method of the PRF, which was compared with the temperature measured using an optical fiber. SPSS software version 22.0 was used for data processing, and the independent sample t-test was used for comparative analysis. P < 0.05 indicated statistical significance. The maximum error between the MRI temperature measurement and the standard value was 3.61°C, whereas the minimum and average errors were 0.01°C and 1.19°C ± 0.78°C, respectively. The temperature measurement technology, which is based on the PRF phase method, has good accuracy in MRI-guided MWA.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.