Abstract
In this paper, the effects of dose and dose rate on the light output of LaBr3:Ce scintillator are studied in detail. The relative light output of the crystal is measured in real time by measuring the photoelectric current of the detector during irradiation at a 137Cs irradiation facility. It is found that the light output of the LaBr3:Ce scintillator continues to decrease with the increase in dose. Permanent radiation damage in LaBr3:Ce is also observed at a cumulative dose of 100kGy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.