Abstract

When a fluid-filled flexible tube is connected to tubing of different impedance, a net flow in either direction can be induced by periodically pinching the flexible section asymmetrically from the ends. We have experimentally demonstrated a variety of conditions under which pumping occurs; including changes in actuator position, size and pinching frequency, transmural pressure, systemic resistance and materials. Data collected includes dynamic pressure and flow-rate measurements at the inlet and outlet of the pump and ultrasound imaging of the tube walls. The net flow rate is highly sensitive to pinching frequency. The pump does not require a closed loop and can sustain a pressure head. We have also shown that a flexible, yet inelastic material is a sufficient condition for impedance-driven flow. A micro-scaled version of the pump was simultaneously tested demonstrating the feasibility of a miniature design.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.