Abstract

The commercially available mechanical vascular anastomotic device with ring-pin technique employs polyethylene rings and stainless steel pins, but even after healing of the anastomotic site, the ring remains. To resolve this disadvantage, the ring material was replaced by biodegradable material, and a ring coupling device was fabricated. This new ring was used to anastomose the femoral vein of rabbits, and morphological observations were made macroscopically and by light and scanning electron microscopy. Of the 58 anastomosed veins, 54 were patent, giving a patency rate of 92.9%. The ring was largely absorbed 15 weeks after anastomosis and completely after 30 weeks. Histologically, re-endothelialisation was observed in 1 week. Infiltration of inflammatory cells was significant around the ring at 3 weeks, but inflammation subsided within 6 weeks. There were no findings indicating adverse effects of the stainless steel pins remaining outside the vascular wall.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.