Abstract

Electroosmotic flow dynamics during a field-amplified sample stacking experiment have been studied experimentally using the periodic photobleaching of a dilute, neutral fluorophore added to the separation buffer. The effects of hydrodynamically injecting different sample plug lengths containing a mixture of arsenic compounds dissolved in 0.125 mM (120, 240, and 600 s) and 41.7 microM (27, 45, and 74 s) phosphate buffer with a separation buffer concentration of 12.5 mM phosphate buffer were examined. Changes in electroosmotic flow during sample stacking and separation were monitored at a rate of 1 Hz. The observed effects of increasing the sample plug length on electroosmotic flow and electrophoretic current agreed qualitatively with predictions by theoretical models presented in the literature. Electroosmotic flow changes on the order of 100% (1.6-3.3 mm/s) were observed. Broadening of the flow monitoring peaks has been used to examine parabolic flow due to the discontinuous buffer systems used for sample stacking.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.