Laser ablation biologic tissue is an important performance of laser application. Based on the combination function of two-wavelength laser to biologic tissue ablation, the semi-continuous two-wavelength laser at 532nm and 1064nm ablating blood tissue was investigated by experiment. The fresh pork and pig liver specimens were ablated at different laser power density and different laser power proportion in water conditions and ablation depth and thermo-damage thickness were measured. The experiment results indicate that there is no effective ablation but deeper thermo-damage of the 1064 nm laser to biologic tissue in water conditions when the 532nm laser power is little but the ablation efficiency sharp inc- reases and the thermo-damage sharp decreases when the 532nm laser power arrives to a threshold. The combination function of two wavelength laser to tissue ablation is appeared here. It indicates that the ablation efficiency and hemostatic function in laser surgery can be adjusted by the two-wavelength laser power proportion.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.