Abstract
The basic principle of tempo-spatially mixed modulated Fourier transform imaging spectrometer (FTIS) based on savart plates is outlined. A calibration method of pixel response non-uniformity of charge-coupled device (CCD) camera in such type of instrument is presented. The method which uses column-flat-fields can avoid the influence of interference fringes. The use of polychromatic calibration source can solve the problem of the slant of the fringes in large optical path difference areas. The procedure of calibration experiment and the algorithm of data processing are detailed described. Two groups of relative radiometric calibration coefficient are obtained through the method of least-square. The original images are corrected by using the coefficients to validate its calibration effect. The results indicated that the method can obviously improve the uniformity of pixels and the vignetting artifacts and defect of the instrument can be well corrected. This study provides a theoretical guidance for study, design, modulation, experiment and engineering of FTIS.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.