The objective of this study was to measure the pulsatile forces acting on a symmetric, bifurcated endoluminal stent graft to validate a computational fluid dynamics (CFD) and analytic model so that they can be used for various graft dimensions. We used a load cell to measure the force owing to the movement of an acrylic model of a bifurcated stent graft under pulsatile flow. This was then simulated with a CFD and analytic model. The main features of the experimental pulsatile force data and the CFD results were consistent. The results showed that the total force was proportional to the inlet pressure cycle. The force rose from 3.32 N at 130 mm Hg systolic to 17.5 N at 250 mm Hg systolic pressure. For the more variable regions of the flow, the experimentally measured forces lagged the computational and analytic results. The CFD and analytic models provide approximate descriptions for the forces acting on a bifurcated stent graft subjected to pulsatile flow. Such models should be of assistance to designers of endoluminal stent grafts.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.