Abstract

In this paper we present the results of the ion mobility measurements made in pure carbon tetrafluoride (CF4) and gaseous mixtures of argon with carbon tetrafluoride (Ar-CF4) for pressures ranging from 6 to 10 Torr (8–10.6 mbar) and for low reduced electric fields in the 10 Td to 25 Td range (2.4-6.1 kV⋅cm−1⋅bar−1), at room temperature. The time of arrival spectra revealed only one peak throughout the entire range studied which was attributed to CF3+. However, for Ar concentrations above 70%, a bump starts to appear at the left side of the main peak for reduced electric fields higher than 15 Td, which was attributed to impurities. The reduced mobilities obtained from the peak centroid of the time-of-arrival spectra are presented for Ar concentrations in the 5%–95% range.

Highlights

  • : In this paper we present the results of the ion mobility measurements made in pure carbon tetrafluoride (CF4) and gaseous mixtures of argon with carbon tetrafluoride (Ar-CF4) for pressures ranging from 6 to 10 Torr (8–10.6 mbar) and for low reduced electric fields in the 10 Td to 25 Td range (2.4-6.1 kV·cm−1·bar−1), at room temperature

  • For Ar concentrations above 70%, a bump starts to appear at the left side of the main peak for reduced electric fields higher than 15 Td, which was attributed to impurities

  • The range of the reduced electric field values used to determine the ions’ mobility is limited by two factors: one is the occurrence of electric discharges at high E/N, and the other is the deterioration of the time of arrival spectra for very low values of E/N, which has been attributed to collisions between the ions and impurity molecules

Read more

Summary

Introduction

: In this paper we present the results of the ion mobility measurements made in pure carbon tetrafluoride (CF4) and gaseous mixtures of argon with carbon tetrafluoride (Ar-CF4) for pressures ranging from 6 to 10 Torr (8–10.6 mbar) and for low reduced electric fields in the 10 Td to 25 Td range (2.4-6.1 kV·cm−1·bar−1), at room temperature. The time of arrival spectra revealed only one peak throughout the entire range studied which was attributed to CF+3. For Ar concentrations above 70%, a bump starts to appear at the left side of the main peak for reduced electric fields higher than 15 Td, which was attributed to impurities. The reduced mobilities obtained from the peak centroid of the time-of-arrival spectra are presented for Ar concentrations in the 5%–95% range.

Methods
Results
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.