Abstract

BackgroundManual skull drilling is an old but in modern neurosurgery still established procedure which can be applied quickly and universally in emergency situations. Electrical drilling requires more complex equipment and is usually reserved to the Operating Room (OR). It also seems desirable to apply an electrical drill for bedside usage but a suitable product does not exist so far.MethodOur experimental study using a manually and an electrically driven skull drill included a total of 40 holes drilled into synthetic biomechanical sheets. Half of the holes were produced with a prototype electrical drilling machine of the company Kaiser Technology and half of them with a traditional manual drill. Different drilling parameters such as the geometry of the borehole, the drilling forces and the drilling vibrations were captured during all experiments.ResultsThe electrical drilling needed higher vertical force by the operators and a longer time to penetrate the sheet. A reason was the relatively lower rotational speed provided by this particular drill. When drilling electrically the vibrations were substantially less which in turn led to a more precise shape of the holes (revealed by observation via a microscope).ConclusionsThe electrification of bedside drilling can in principle enable emergency craniostomies to be performed with greater ease and accuracy. The power of the electric drive, however, must be at least equivalent to the power of the traditional manual drill. Otherwise, the vertical forces exerted on the scull by the operator become inhibitive. The challenge is to combine cost-efficiency and re-sterilizability of an electrically driven drilling machine which at the same time is small and simple enough to qualify for emergency applications.

Highlights

  • Skull drilling is a surgical procedure that has been practiced since prehistoric times [1, 2, 3, 4]

  • The electrification of bedside drilling can in principle enable emergency craniostomies to be performed with greater ease and accuracy

  • The challenge is to combine cost-efficiency and re-sterilizability of an electrically driven drilling machine which at the same time is small and simple enough to qualify for emergency applications

Read more

Summary

Introduction

Skull drilling is a surgical procedure that has been practiced since prehistoric times [1, 2, 3, 4]. Some technical modifications to reduce these risks have been developed, for instance a pre-adjustable distance holder [1] or a percutaneous needle which replaces the turning drilling bit [14, 15]. Prediction and control of thrust forces and torques as well as possibly an optimized drill design are typical tasks related to mechanical engineering; see e.g. the publications [16, 17, 18]. This contribution tries to bridge the gap between engineering work and every days neurosurgical practice. It seems desirable to apply an electrical drill for bedside usage but a suitable product does not exist so far

Methods
Discussion