Abstract
The paper presents the results of an experimental activity focused on the measurement of electromagnetic fields generated by magnetic resonance imaging systems, carried out with the purpose of assessing exposure of workers to static, low frequency and radio-frequency electromagnetic fields , as required by the EU Directive 2004/40/EC. The main contribution of the paper is that it presents a comparative study of exposure carried out during operation of the MRI system and close to the sources, with different scanning sequences and RF coils. Results show that the static magnetic field follows the theoretical values only close to the magnet due to the effect of the walls, with a few percent variation due to the operating conditions; exposure to the magnetic field in the low frequency band is mainly determined by emissions generated by the power grid at 50 Hz and by the pulses driving the gradient coils; and finally the radio-frequency band only shows contributions by the diagnostic pulses at the resonance frequency, with exposure levels usually below the limits contained in the Directive.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.