Abstract

In this work, we studied experimentally the influence of x-ray spectrum on the contrast-to-noise ratio (CNR) and the average glandular dose (MDG) for two digital mammography systems: Senographe 2000D (GE Medical Systems) and Lorad Selenia (Hologic), with indirect and direct detector imaging technology, respectively. CNR and MGD were determined using PMMA phantoms simulating breasts with thicknesses of 4 cm and 6 cm. All available anode/filter combinations of the systems were evaluated for a wide range of tube voltages values. Results indicated that the Rh/Rh combination provides the highest image quality with the lower mean glandular dose for the Senographe 2000D system. For the Lorad Selenia system, the W/Ag combination at 30 kV showed the best performance, in terms of dose saving and image quality improvement in relation to all tube voltage range. The comparison between the optimal x-ray spectra and those selected by the AEC mode showed that this automatic selection mechanism could be readjusted to optimize the relationship between image quality and dose.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.