Abstract

PurposesThe introduction of digital breast tomosynthesis (DBT) into the French breast cancer screening program is forecast by the authorities. The aim of the present study was to evaluate image quality phantoms to be used as internal quality controls. MethodsSeven breast phantoms dedicated to quality control in mammography were evaluated on reconstructed DBT images: ACR Model 015, BR3D, DBT QC model 021, Mam/Digi-EPQC, MTM100, TOMOMAM® and TOMOPHAN®. Two representative image parameters of DBT images were studied: image score and z-resolution, when inserts were included in the phantom, on five DBT systems of three different brands. Three observers were involved. ResultsThe MTM100, Mam/Digi-EPQC, BR3D, DBT QC model 021 phantoms' images presented artefacts affecting the image score. The ACR Model 015, TOMOMAM® and TOMOPHAN® phantoms appeared to be pertinent for DBT image score analysis. Due to saturation artefacts, Z-resolution results were not coherent with the theory for all phantoms except by using aluminium beads in the TOMOMAM® phantom. ConclusionsPhantom manufacturers should be encouraged to collaborate with DBT system manufacturers in order to design universal phantoms suitable for all systems for more complete quality control. From our study we can propose several specifications for an ideal and universal phantom designed for internal quality control in DBT. Phantoms should allow sensitive image score measurements. The background structure should be realistic to avoid artefacts. Phantoms should have a standard breast-like shape and size.